Liquid Chromatography:
Ranges:
0.01.. 999.99 mS/cm
Accuracy:
< 2 % from FS (0.01-300ms/cm Range)
Flowpath ID 0.25/0.75 mm PEEK, max. pressure 135 bar, connection port 10-32
Interface: Analog 0-1V
© Innovative Sensor Technologies GmbH 2015
info@istech.at
It can be used to replace existing standard conductivity detectors.
The standard range is 0..10000µS/cm, for eluent monitoring 0..999.99mS/cm.
HPLC/IC specifications ( 0..10 ml/min):
Monitor specifications ( 0..10/100 ml/min):
Ranges:
0..100, 0..500,1000, 0..10000 µS/cm
Accuracy:
< 2 % from FS
Deviation:
< 0.5 % from FS
Noise level (wet) Range 100µS/cm: <0.2 nS/cm Background 10µS/cm
Noise level (wet) Range 500µS/cm: < 4 nS/cm Background 200µS/cm
Drift(Temperature):
<0.006 % of FS /°C
Drift(Time):
<0.006 % of FS / h
Flowpath ID 0.25mm PEEK, max. pressure 135 bar, connection port 10-32
Interface: Analog 0-1V
The complete flowpath is
in PEEK !